Gain Switching for a Detection System to Accommodate a Newly Developed MALDI-Based Quantification Method

DC Field Value Language
dc.contributor.authorSung Hee Ahn-
dc.contributor.authorTaeghwan Hyeon-
dc.contributor.authorMyung Soo Kim-
dc.contributor.authorJeong Hee Moon-
dc.date.available2018-01-05T05:50:59Z-
dc.date.created2017-09-25-
dc.date.issued2017-09-
dc.identifier.issn1044-0305-
dc.identifier.urihttps://pr.ibs.re.kr/handle/8788114/4162-
dc.description.abstractIn matrix-assisted laser desorption ionization time-of-flight mass spectrometry (MALDI-TOF), matrix-derived ions are routinely deflected away to avoid problems with ion detection. This, however, limits the use of a quantification method that utilizes the analyte-to-matrix ion abundance ratio. In this work, we will show that it is possible to measure this ratio by a minor instrumental modification of a simple form of MALDI-TOF. This involves detector gain switching. [Figure not available: see fulltext.]. © 2017, American Society for Mass Spectrometry-
dc.description.uri1-
dc.language영어-
dc.publisherSPRINGER-
dc.subjectDetector gain switching-
dc.subjectMALDI quantification-
dc.subjectMALDI-TOF instrument-
dc.titleGain Switching for a Detection System to Accommodate a Newly Developed MALDI-Based Quantification Method-
dc.typeArticle-
dc.type.rimsART-
dc.identifier.wosid000407776200027-
dc.identifier.scopusid2-s2.0-85027414267-
dc.identifier.rimsid60319ko
dc.date.tcdate2018-10-01-
dc.contributor.affiliatedAuthorTaeghwan Hyeon-
dc.contributor.affiliatedAuthorMyung Soo Kim-
dc.identifier.doi10.1007/s13361-017-1711-2-
dc.identifier.bibliographicCitationJOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, v.28, no.9, pp.1987 - 1990-
dc.citation.titleJOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY-
dc.citation.volume28-
dc.citation.number9-
dc.citation.startPage1987-
dc.citation.endPage1990-
dc.date.scptcdate2018-10-01-
dc.description.wostc1-
dc.description.scptc1-
dc.description.journalClass1-
dc.description.journalRegisteredClassscie-
dc.description.journalRegisteredClassscopus-
dc.subject.keywordAuthorDetector gain switching-
dc.subject.keywordAuthorMALDI quantification-
dc.subject.keywordAuthorMALDI-TOF instrument-
Appears in Collections:
Center for Nanoparticle Research(나노입자 연구단) > 1. Journal Papers (저널논문)
Files in This Item:
08.gain switching.pdfDownload

qrcode

  • facebook

    twitter

  • Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.
해당 아이템을 이메일로 공유하기 원하시면 인증을 거치시기 바랍니다.

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse