BROWSE

Related Scientist

cho,yoonkyoung's photo.

cho,yoonkyoung
인공지능및로봇기반합성연구단
more info

ITEM VIEW & DOWNLOAD

Electrospun TiO2 nanofiber integrated lab-on-a-disc for ultrasensitive protein detection from whole blood

Cited 24 time in webofscience Cited 29 time in scopus
1,544 Viewed 328 Downloaded
Title
Electrospun TiO2 nanofiber integrated lab-on-a-disc for ultrasensitive protein detection from whole blood
Author(s)
Won Seok Lee; Vijaya Sunkara; Ja-Ryoung Han; Yang-Seok Park; Yoon-Kyoung Cho
Publication Date
2015-01
Journal
LAB ON A CHIP, v.15, no.2, pp.478 - 485
Publisher
ROYAL SOC CHEMISTRY
Abstract
ELISA-based devices are promising tools for the detection of low abundant proteins in biological samples. Reductions of the sample volume and assay time as well as full automation are required for their potential use in point-of-care diagnostic applications. Here, we present a highly efficient lab-on-a-disc composed of a TiO2 nanofibrous mat for sensitive detection of serum proteins with a broad dynamic range, with only 10 μL of whole blood within 30 min. The TiO2 nanofibers provide high specific surface area as well as active functional groups to capture large amounts of antibodies on the surface. In addition, the device offers efficient mixing and washing for improving the signal-to-noise ratio, thus enhancing the overall detection sensitivity. We employ the device for the detection of cardiac biomarkers, C-reactive protein (CRP) and cardiac troponin I (cTnI), spiked in phosphate-buffered saline (PBS) as well as in serum or whole blood. The device exhibited a wide dynamic range of six orders of magnitude from 1 pg mL−1 (~8 fM) to 100 ng mL−1 (~0.8 pM) and a low detection limit of 0.8 pg mL−1 (~6 fM) for CRP spiked in CRP-free serum and a dynamic range of 10 pg mL−1 (~0.4 pM) to 100 ng mL−1 (~4 nM) with a detection limit of 37 pg mL−1 (~1.5 pM) for cTnI spiked in whole blood.
URI
https://pr.ibs.re.kr/handle/8788114/2155
DOI
10.1039/c4lc00900b
ISSN
1473-0197
Appears in Collections:
Center for Soft and Living Matter(첨단연성물질 연구단) > 1. Journal Papers (저널논문)
Files in This Item:
Electrospun Tio2 nanofiber integrated lab on a disc for ultrasensitive protein detection from whole blood.pdfDownload

qrcode

  • facebook

    twitter

  • Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.
해당 아이템을 이메일로 공유하기 원하시면 인증을 거치시기 바랍니다.

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse