BROWSE

Related Scientist

cnir's photo.

cnir
뇌과학이미징연구단
more info

ITEM VIEW & DOWNLOAD

Identification and Reduction of Image Artifacts in Non–Contrast-Enhanced Velocity-Selective Peripheral Angiography at 3T

Cited 10 time in webofscience Cited 10 time in scopus
1,426 Viewed 340 Downloaded
Title
Identification and Reduction of Image Artifacts in Non–Contrast-Enhanced Velocity-Selective Peripheral Angiography at 3T
Author(s)
Taehoon Shin; Qin, Qin; Jang-Yeon Park; Crawford, Robert S.; Rajagopalan, Sanjay
Subject
velocity-selective magnetization preparation, ; noncontrast-enhanced MR angiography at 3T, ; peripheral angiography
Publication Date
2016-08
Journal
MAGNETIC RESONANCE IN MEDICINE, v.76, no.2, pp.466 - 477
Publisher
WILEY-BLACKWELL
Abstract
Purpose: To identify and reduce image artifacts in non–contrast-enhanced velocity-selective (VS) magnetization-prepared peripheral MR angiography (MRA) at 3T. Methods: To avoid signal loss in the arteries, double and quadruple refocused VS excitation pulse sequences were designed that were robust to a wide range of B0 and B1 offset. To suppress stripe artifact and background signal variation, we successively applied two VS preparations with excitation profiles shifted by half the period of the stripes. VS-MRA using single, double, and quadruple refocused VS preparations was tested in healthy subjects and a patient. Results: In the regions of large B0 and B1 offsets, arterial signal loss was yielded by single refocused VS preparation, but was avoided with double or quadruple refocused preparations. Compared with single VS preparation, the two consecutive preparations with shifted excitation profiles substantially reduced the stripe artifact and background signal variation, as demonstrated by increased mean and decreased standard deviation of relative contrast-to-noise ratio. The proposed VS-MRA identified multilevel disease in the femoral arteries of the patient, as validated by digital subtraction angiography. Conclusion: Two multiple refocused VS magnetization preparations with shifted excitation profiles yield artifact-free peripheral angiograms at 3T. Magn Reson Med 76:466–477, 2016. © 2015 Wiley Periodicals, Inc. © 2015 Wiley Periodicals, Inc
URI
https://pr.ibs.re.kr/handle/8788114/2736
DOI
10.1002/mrm.25870
ISSN
0740-3194
Appears in Collections:
Center for Neuroscience Imaging Research (뇌과학 이미징 연구단) > 1. Journal Papers (저널논문)
Files in This Item:
16.Identification and Reduction of Image Artifacts.pdfDownload

qrcode

  • facebook

    twitter

  • Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.
해당 아이템을 이메일로 공유하기 원하시면 인증을 거치시기 바랍니다.

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Browse